A minimum ISO 4 air quality in the work area
Patent of non leakage
Every ULPA filters will be tested to ensure non leakage.
Add to Cart
Safe and Reliable, Energy-Efficiency
DC fan and LED light make the operation power consumption save 30%.
Intelligent auto-supply system of imported blower ensures the change of air volume less than 10% when the resistance of filter increases 50% and enhance safety.
The alarm system for safe operation height is interlocked with light. At the same time light is interlocked with sterilizing.
Leak detection in cabinet, no leakage at 500Pa.
UV light with set timer, safer and high efficiency.
Specification of Laboratory Furniture
Exterior size (WXDXH) | 1200X795X2050 | 1500X795X2050 | 1800X795X2050 | 2000X795X2050 |
Filtration efficiency | ULPA: more than 99.995%, @0.12µm | |||
Inflow velocity | 0.55 m/s | |||
Downflow velocity | 0.35 m/s | |||
Vibration | Less than 5µm, center of countertop | |||
Opening height | 200mm | |||
Personnel protection | A. Total colony in impaction sampler ≤10 CFU/ Time | |||
B. Total colony in impaction sampler ≤5 CFU/ Time | ||||
Product protection | Total colony in culture dish ≤5 CFU/ Time | |||
Cross-contamination protection | Total colony in culture dish ≤2 CFU/ Time |